Concentration from A260
Nucleic acids absorb strongly at 260 nm. The Beer–Lambert law makes absorbance proportional to concentration, and for routine work the molar absorptivity is folded into a conventional conversion factor: the concentration that gives an absorbance of 1.0 in a 1 cm path.
Concentrationc[μg/mL=ng/μL]=lA260×F×D - A260
- absorbance at 260 nm
- F
- conversion factor: 50 µg/mL for dsDNA, 33 for ssDNA, 40 for RNA
- D
- dilution factor of the sample you measured (1 if undiluted)
- l
- path length in cm (NanoDrop values are already normalised to 1 cm)
Purity is judged from absorbance ratios: A260/A280≈1.8 for DNA and ≈2.0 for RNA (protein and phenol absorb near 280 nm), and A260/A230≈2.0–2.2 (salts, guanidine and carbohydrates absorb near 230 nm). The factors assume a long, mixed-sequence polymer; short oligos need a sequence-specific ε.
References. Sambrook J, Russell DW. Molecular Cloning: A Laboratory Manual, 3rd ed., Appendix 8. Cold Spring Harbor Laboratory Press, 2001.
Molecular weight, molarity and copy number
Molarity needs the molecular weight, which for a sequence of unknown composition is estimated from its length using average nucleotide masses. These are the constants NEB’s NEBioCalculator uses:
Molecular weight from lengthdsDNA:ssDNA:RNA:MW=Nbp×617.96+36.04MW=Nnt×308.97+18.02MW=Nnt×321.47+18.02[g/mol] A concentration of 1 ng/µL is 1 mg/L, so dividing by the molecular weight and scaling to nanomolar gives:
Mass to molarc[nM]=MWc[ng/μL]×106 Copy numberμLcopies=c[nM]×10−9Lmol×10−6μLL×NA,NA=6.02214076×1023 mol−1 Totals in a volume V (µL) are m=c[ng/μL]×V and n[pmol]=c[nM]×V×10−3. The commonly quoted 650 g/mol per base pair is the sodium salt and gives molarities about 5 % lower.
References. New England Biolabs, NEBioCalculator: dsDNA, ssDNA and RNA mass ↔ moles conversions. Avogadro constant: CODATA 2018 / SI 2019 exact value.
Mass for a target copy number
For qPCR standards and template input, invert the copy-number equation: the mass that holds a given number of copies, and the working concentration that delivers it in the template volume you pipette.
Mass from copiesm[ng]=NAcopies×MW×109,cwork=Vtemplatem,dilution=cworkcstock For example one haploid human genome (≈3.1 × 10⁹ bp) weighs about 3.2 pg, and 10⁶ copies of a 3,000 bp plasmid weigh about 3.1 pg.
Ligation insert:vector molar ratio
Ligation efficiency depends on the number of DNA ends, so the insert is dosed as a molar ratio to the vector. Because both are dsDNA, the molecular-weight constants cancel and only the length ratio matters:
Insert massminsert=mvector×LvectorLinsert×vectorinsert Typical ratios are 1:1 to 3:1 for sticky-end cloning and 3:1 to 5:1 for blunt ends or short inserts; very short inserts (under ~200 bp) often need higher ratios.
Oligonucleotide molecular weight and ε260
For a synthetic oligo the exact base composition is known. The molecular weight is the sum of the anhydrous nucleotide masses, corrected for the missing 5′ phosphate of an unmodified oligo (5′-OH, 3′-OH):
Molecular weightMW=313.21nA+289.18nC+329.21nG+304.19nT−61.96 The extinction coefficient is estimated as the sum of the free nucleotides’ absorptivities, and one OD260 unit (1 mL at A260=1) contains 106/ε nmol:
Base-sum extinction coefficientε260=15400nA+7400nC+11500nG+8700nT[M−1cm−1] Stacking between neighbouring bases lowers the real absorbance (hypochromicity), so nearest-neighbour ε values, as printed on synthesis spec sheets, are lower than this estimate. Molecular weights are checked against Biopython’s molecular_weight.
References. Biopython Bio.SeqUtils.molecular_weight (IUPAC average masses). ε values: the standard molar absorptivities of the deoxynucleoside 5′-monophosphates at 260 nm, neutral pH.